CAN/LIN Triggering and Serial Decode 22
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 339
 
CAN Serial Decode
To set up the oscilloscope to capture CAN signals, see "Setup for CAN 
Signals" on page 335.
To set up CAN serial decode:
1 Press  [Serial] to display the Serial Decode Menu.
2 If the decode line does not appear on the display, press the [Serial] key 
to turn it on.
3 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and 
decode data.
You can use the horizontal Zoom window for easier navigation of the 
decoded data.
See Also • "Interpreting CAN Decode" on page 340
• "CAN Totalizer" on page 341
• "Interpreting CAN Lister Data" on page 342
• "Searching for CAN Data in the Lister" on page 343
To display CAN serial decode, see "CAN Serial Decode" on page 339.
For CAN triggering set up see "CAN Triggering" on page 337.
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the 
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode 
softkey to set the trigger mode from Auto to Normal.