198 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
12 Acquisition Control
 
When capturing multiple infrequent trigger events it is advantageous to 
divide the oscilloscope's memory into segments. This lets you capture 
signal activity without capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO 
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see 
"Measurements, Statistics, and Infinite Persistence with Segmented 
Memory" on page 199) to show infinite persistence across all acquired 
segments. See also "To set or clear persistence" on page 131 for details.
To acquire to
segmented
memory
1 Set up a trigger condition. (See Chapter 10, “Triggers,” starting on page 
141 for details.)
2 Press the [Acquire] key in the Waveform section of the front panel.
3 Press the Segmented softkey.
4 In the Segmented Memory Menu, press the Segmented softkey to enable 
segmented memory acquisitions.
5 Press the #ofSegs softkey and turn the Entry knob to select the 
number of segments into which you would like to divide the 
oscilloscope's memory.
Memory can be divided into as few as two segments and as many as 
1000 segments, depending on the oscilloscope model.
6 Press the [Run] or [Single] key.
The oscilloscope runs and fills a memory segment for each trigger event. 
When the oscilloscope is busy acquiring multiple segments, the progress is 
displayed in the upper right area of the display. The oscilloscope 
continues to trigger until memory is filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity, 
consider selecting the Normal trigger mode to prevent AutoTriggering. See 
"To select the Auto or Normal trigger mode" on page 178.