364 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24 I2C/SPI Triggering and Serial Decode
Frame (Start: Addr7: Read: Ack: Data) or Frame (Start: Addr7: Write: Ack: Data)
The oscilloscope triggers on a read or write frame in 7- bit addressing
mode on the 17th clock edge if all bits in the pattern match. For
triggering purposes, a restart is treated as a start condition.
Frame (Start: Addr7: Read: Ack: Data: Ack: Data2) or Frame (Start: Addr7: Write:
Ack: Data: Ack: Data2)— The oscilloscope triggers on a read or write
frame in 7- bit addressing mode on the 26th clock edge if all bits in
the pattern match. For triggering purposes, a restart is treated as a
start condition.
10-bit Write — The oscilloscope triggers on a 10- bit write frame on the
26th clock edge if all bits in the pattern match. The frame is in the
format:
Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data)
For triggering purposes, a restart is treated as a start condition.
SDA
SCL
Control
byte
Read
R/
Ack
Data
Ack
Trigger point
Start or
Restart
Condition
SDA
SCL
Stop
Condition
Start or
Restart
Condition
Trigger point
17th clock edge
Address
Read
Write
R/
Ack
Data
Ack
Read
Write
Start
Condition
Address R/W Ack
Data
Ack
Data 2
Trigger point
26th clock edge
Ack
Stop
Condition

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals