386 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
25 I2S Triggering and Serial Decode
 
If you select an "Armed" value that is equal to or greater than the 
"Trigger" value, the "Trigger" value will be increased so that it is 
always greater than the "Armed" value.
• Decreasing value — similar to the description above except the trigger 
occurs on a descreasing data word value, and the "Armed" value is 
the value to which the data must rise in order to re- arm the trigger.
6 Press the Base softkey and select a number base for entering data 
values:
• Binary (2's complement).
When Binary is selected, the Bits softkey appears. This softkey opens 
the I2S Bits Menu for entering data values.
When the trigger qualifier requires a pair of values (as with In 
Range, Out of Range, Increasing value, or Decreasing value), the first 
softkey in the I2S Bits Menu lets you select which value of the pair.
In the I2S Bits Menu, press the Bit softkey and rotate the Entry knob 
to select each bit; then, use the 01X softkey to set each bit value to 
zero, one, or don't care. You can use the Set all Bits softkey to set all 
bits to the value chosen on the 01X softkey.
• Signed decimal.
When Decimal is selected, the softkey(s) to the right let you enter 
decimal values with the Entry knob. These softkeys can be Data, <, >, 
or  Threshold depending on the selected trigger qualifier.
"Trigger" value
Trigger
Trigger
No trigger
"Armed" value
If the setup does not produce a stable trigger, the I2S signal may be slow enough that the 
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode 
softkey to set the trigger mode from Auto to Normal.