412 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
27 UART/RS232 Triggering and Serial Decode
When #Bits is set to 9 in the UART/RS232 Bus Configuration Menu,
the 9th (alert) bit is displayed directly to the left of the ASCII value
(which is derived from the lower 8 bits).
4 Optional: Press the Framing softkey and select a value. In the decode
display, the chosen value will be displayed in light blue. However, if a
parity error occurs the data will be displayed in red.
5 If the decode line does not appear on the display, press the [Serial] key
to turn it on.
6 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
decode data.
You can use the horizontal Zoom window for easier navigation of the
acquired data.
See Also "Interpreting UART/RS232 Decode" on page 413
"UART/RS232 Totalizer" on page 414
"Interpreting UART/RS232 Lister Data" on page 415
"Searching for UART/RS232 Data in the Lister" on page 415
NOTE
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow
enough that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then
press the Mode softkey to set the trigger mode from Auto to Normal.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals