112 PL-GPC 220 User Manual
9 Appendix
General Safety Information
As part of the electrical safety regulations, forming part of the overall CE
compliance, the unit is fitted with a surge safe device located at the rear of the
instrument. This device protects the instrument from high voltage line surges
by filtering the incoming supply.
Solvent vapors
Liberal use of organic solvents in close proximity to the instrument may activate the
vapor sensors, causing the instrument to shutdown.
➔ Exercise care when using solvents close to the instrument.
Insulation resistance test fails
The surge safe device causes the instrument to fail the insulation resistance test of the
standard portable appliance test (PAT Test).
➔ Unplug the surge safe device before the PAT test is performed.