74 PL-GPC 220 User Manual
7 Maintenance
Decontamination
Decontamination
1 Excess quantities of solvent spilt on or inside the instrument should be
mopped up using absorbent cloths, followed by repeated wiping down with
soft cloths moistened with acetone until the last traces of the hazardous
liquid have been removed.
2 Identify and, if possible, correct the source of the leak.
3 Inspect cabling, parts and surfaces to determine whether or not any
damage has occurred.
4 Allow the the instrument to dry completely before reconnecting power.
5 Dispose of contaminated waste appropriately.
WARNING
Intoxication
Potentially hazardous liquid spilt in the instrument.
Wear appropriate personal protective equipment (gloves, safety glasses, lab coat
and respirator if the level of hazard has been risk-assessed to be sufficiently high).
NOTE
To be carried out only by trained service personnel.