EasyManua.ls Logo

Agilent Technologies PL-GPC 220 - Modes of Operation; Pump Flow Control; RI Detector

Agilent Technologies PL-GPC 220
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PL-GPC 220 User Manual 33
Using the Module
4
The Graphical User Interface
Modes of Operation
The PL-GPC 220 will be in one of the following operation modes:
STANDBY – indicated by a RED LED on both the interface and the
instrument display panel is the boot-up mode, in which all modules are off.
This is the mode of operation the PL-GPC 220 is set for routine
maintenance, such as solvent or column exchange or in any circumstance
when the oven needs to be opened.
IDLE – indicated by an AMBER LED. Use this mode to control the
instrument in order to maintain oven temperature and a low flow rate
(configurable, but typically set to 0.1 mL/min) so that the system is ticking
over but not running samples. The carousel heaters are also turned off by
default in IDLE mode.
RUN – indicated by a GREEN LED, is the fully functional, operational mode.
The oven temperature is at set point, the flow rate is typically at
1.0 mL/min, the carousel heaters are on, and the instrument is ready to run
samples.
Following boot-up, the instrument will typically default to the STANDBY
mode, unless the instrument is switched off in either the IDLE or RUN mode,
in which case it will re-boot back into the mode it was switched off in.
Pump Flow Control
The pump flow can be manually switched on/off in two places: the first using
the on/off selection button in the pump module box, and the second is located
in the Manual Operation tab. Both options allow the pump flow to be
stopped/started in all three modes of operation.
RI Detector
•Autozero – initializing the RI autozero function is a procedure, which should
be performed after a purge cycle.
•Purge – open purge valve on RI reference cell (the length of RI purge is set in
the RI configuration window). This procedure will de-stabilize the system; it
is recommended that the baseline be monitored and if excessive drift is
noticed then another purge cycle is performed. This procedure should be
repeated until a stable baseline is observed.

Table of Contents

Related product manuals