EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>PL-GPC 220

Agilent Technologies PL-GPC 220 User Manual

Agilent Technologies PL-GPC 220
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
94 PL-GPC 220 User Manual
7 Maintenance
Differential Refractive Index Detector
Purge the DRI Cell
The following problems are usually indicative that the reference cell in the DRI
detector requires purging with fresh solvent:
Drift on the baseline
Excessive noise on the baseline
Lack of sensitivity
1 In RUN mode open the purge valve by pressing the PURGE button.
2 While in PURGE mode, enter the Pump configuration and increase the flow
to 2.0 mL/min (if using a purge column).
3 Allow the system to purge at 2.0 mL/min for 15 min, during which time the
baseline can be monitored and should give a flat noise free baseline. If the
baseline is off scale, press the Autozero button on the front panel.
4 After approximately 15 min, reduce the FLOW back to the required
operating flow (for example 1.0 mL/min) and wait for the pump flow rate to
reach the set value.
5 Press the Purge button to close the purge valve.
6 Allow the system adequate time to settle after the purge.
For high temperature applications, this settling time can be up to 8 hours.
NOTE
At this point the PURGE light will go out and the Purge button will have to be pressed again
once the RI has zeroed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PL-GPC 220 and is the answer not in the manual?

Agilent Technologies PL-GPC 220 Specifications

General IconGeneral
BrandAgilent Technologies
ModelPL-GPC 220
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals