EasyManua.ls Logo

Agilent Technologies PL-GPC 220 - Performance Specifications

Agilent Technologies PL-GPC 220
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 PL-GPC 220 User Manual
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Tab le 2 Performance Specifications of the PL-GPC 220
Unit Type Specification
Pump Flow Range 0.2 – 10 mL/min Pump
0.1 – 5.0 mL/min software
controlled
Flow Rate Reproducibility
1
1
The technical specifications have been determined by experimentation and reflect the total system
performance. The values quoted have been calculated from analyses performed in THF at 40 °C and
in TCB at 160 °C.
0.07 % RSD, or 0.02 min SD
whatever is greater, based on
retention time at constant
room temperature
Oven Temperature Range Ambient to 220 ° C
Temperature Stability <0.05 ° C/h
Column Capacity 6 x 30 cm columns
Detector Differential Refractometer Deflection
Cell Volume 8 µL
Noise
1
80 µV
Drift
1
10 mV/h
Wavelength 890 nm
Autosampler Carousel Capacity 40 x 2 mL Vials
Injection Volume flushed fixed loop; 5 – 500 µL
Precision
1
1%
Instrument Control Microprocessor
Instrument Interface Windows 7 Advanced PC
Control