52 PL-GPC 220 User Manual
4 Using the Module
Setting Up the PL-GPC 220
Changing the Injection Loop
1 Change the Idle and Low flow rates to 0.2 mL/min and the Idle temperature
to 40 °C.
2 Put the instrument into IDLE mode and let the oven cool.
Only continue if oven is at 40 °C.
3 Put the instrument into STANDBY mode and open the oven lids.
4 Replace the injection loop with the new loop size.
5 Continue with following procedure: “Running the PL-GPC 220” on page 49.
Preparations The column is in RUN mode.