PL-GPC 220 User Manual 59
Troubleshooting and Diagnostics
5
General Troubleshooting Guide
If none of the above steps occur:
Ensure +5 V is present by observing the rotation of stirrer magnet.
Ensure +12 V is present by observing the rotation of ventilation fans
in the baseplate.
Autosampler Problems
Please contact your local Agilent service representative.
RI Problems
Fails to autozero correctly
Purge the RI thoroughly.
Check the purge column is not blocked (if fitted), by observing a
significant decrease in system pressure when the purge valve opens.
Low sensitivity
Clean the RI cell thoroughly (for example sonicate in 10 % nitric acid).