EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>PL-GPC 220

Agilent Technologies PL-GPC 220 User Manual

Agilent Technologies PL-GPC 220
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
54 PL-GPC 220 User Manual
4 Using the Module
Running Samples on the PL-GPC 220
Loading the Samples
1 Pull out the autosampler carousel drawer.
2 Load the autosampler carousel with the sample and rinse vials in the
appropriate vial positions.
3 Firmly push the autosampler carousel drawer back into position.
You will hear an audible click as the autosampler fits into place.
4 Ensure that the data collection software you are using is ready to collect
and the sequence run list has been defined (remember data will not be
collected for rinse vials).
5 On the autosampler area of the Control software, click the Autosampler
Offline button.
This will now change to Autosampler Online and the sequence will begin. You
can follow the status of the run sequence in the software.
6 At any time, you can stop the autosampler sequence by clicking the
Autosampler Online button.
The software will then prompt you to select a new start vial if the sequence
is restarted.
Preparations Start the log file to monitor the condition of the instrument during operation.
Ensure that there is sufficient eluent to carry out the analysis (you can use the time estimator
tool on the Control software once you have defined the autosampler sequence).
Ensure that there is sufficient capacity in the waste container.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PL-GPC 220 and is the answer not in the manual?

Agilent Technologies PL-GPC 220 Specifications

General IconGeneral
BrandAgilent Technologies
ModelPL-GPC 220
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals