General Safety Information
Precautions
The PL-GPC 220 is designed to control the temperature of the column oven
and autosampler at temperatures up to 220 °C.
Vapor sensors are used in both modules of the PL-GPC 220 to alert the
operator to solvent leaks.
Fire, intoxication and damage to the instrument
Hazardous material on or inside the instrument.
➔ Do not use the instrument if material is split on or inside the instrument.
➔ Carry out appropriate decontamination method recommended by Agilent
Technologies.
➔ If using methods not recommended by Agilent Technologies, check with Agilent
Technologies that the method will not damage the instrument.
➔ Perform periodic leak checks on all wetted components, tubing and joints.
➔ Test the over temperature protection systems for the solvent module, oven and
autosampler on a regular basis.
➔ Do not allow flammable and/or toxic solvents to accumulate.
➔ Follow recommended procedures and protocols for evacuation and disposal of
flammable or toxic solvents. Never dispose of such products through municipal
waste systems.
Hot surfaces
The surface of the inner oven, autosampler and heated waste line will get extremely
hot during normal operation (up to 220 °C).
➔ Handle these components carefully.
➔ Wear suitable protective gloves.