48 PL-GPC 220 User Manual
4 Using the Module
Setting Up the PL-GPC 220
9 If a viscometer is fitted, additional flushing of the viscometer pressure
transducers is required - see the G7821A PL-GPC 220 Viscometer manual.
10 Once the system is totally flushed into the desired chromatography solvent:
a Turn off the pump.
b Close the DRI purge valve.
c Connect in the column set
Before turning on the pump to flush the columns ensure the IDLE flow is set at a rate
<0.3 mL/min.