Testing Vapor Sensors
8 Wait for two minutes with the alarm sounding, then check that the solvent
flow has stopped by watching for drips from the return tube.
9 If the alarm condition does not occur or if the pump flow has not stopped
after three minutes then switch off and disconnect the power cords from
both units.
10 Open the oven and remove the paper. Dispose of it appropriately.
11 Leave the oven open for one minute to allow the acetone vapour to disperse.
12 Close both oven lids.
13 Clear the error on the Control software
This concludes the oven vapour sensor test.