92 PL-GPC 220 User Manual
7 Maintenance
Back Pressure Regulator (BPR)
3 Fit the replacement BPR with the arrow in the direction of the flow, for
example from right to left.
4 Turn on the flow and open the PURGE valve. Check the system
backpressure and check for leaks at the joints disturbed.