Controller / User Interface | 4-29
Default Cell Pres
Displays the default absolute pressure for the Measuring Cell
(mmHg). This value is used in place of the Measuring Cell pressure
when active pressure compensation is disabled. Enter a value to place
the pressure compensation into fixed mode (the approximate Measur-
ing Cell pressure).
Refer to the appropriate settings, below, for jumper selectable pres-
sure ranges for the corresponding 0ā100 PSIA (0ā700 kPaa), 0ā30 PSIA
(0ā210 kPaa), or 0ā500 PSIA (0ā3450 kPaa) Pressure Transducer.
If the 0ā100 PSIA pressure transducer is being used, the jumpers
(JP301/JP302) on the Customer I/O board allow you to re-scale the
pressure range. See ā0ā100 PSIA Pressure Transducerā chart below.
0ā100 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA mmHg JP301 JP302
0ā100 0ā5171.5 A A
0ā33 0ā1706.6 A B
33ā66 1706.6ā3412.2 B B
66ā100 3413.2ā5171.5 C B
While the 0ā30 PSIA pressure transducer can not be re-scaled, its
jumper settings are listed in the ā0ā30 PSIA Pressure Transducerā chart
below.
0ā30 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA Pressure Range JP301 JP302
0ā30 0ā1552 mmHg A A
If the 0ā500 PSIA pressure transducer is being used, the jumpers
(JP301/JP302) on the Customer I/O board allow you to re-scale the
pressure range. The maximum allowable Cell Pressure is 200 psig
(1380 kPag). See ā0ā500 PSIA Pressure Transducerā chart below.
0ā500 PSIA Pressure Transducer Chart (High Pressure/Low Range Kit):
Jumper Settings, Customer I/O Board
PSIA Pressure Range JP301 JP302
0ā166.667 0ā8619.16 mmHg A B
If changing the jumpers, be sure to make corresponding changes in
the S93XS System 200 Configurator Software.
ļ”
NOTE