Maintenance and Troubleshooting | 6-37
Observe the top line of the Display readout. The following messages
are displayed during the Auto-Setup sequence:
• Adjusting PMT Level
• Adjusting Lamp Pulses
• Adjusting PMT Balance
• Adjusting PMT Level
• Adjusting Lamp Pulses
• Setting Up ALC
• Off/Completed
When the message “Off/Completed” is displayed, the Auto-Setup is
complete.
4. From the Optical Bench dialog box, check the PMT Level and PMT
Balance values again. The PMT Level should be 7200–7400 mV, while
the PMT Balance should be 2000–8000 mV.
5. If further adjustments are required, refer to “Manipulating the PMT
Level and PMT Balance,” following this section.
Cancel Auto-Setup
To cancel the Auto-Setup, click Cancel Auto-Setup on the Optical Bench
dialog box.