2-6 | Model 933S UV Analyzer
Sample Stream Compatibility
The analyzer will work in applications beyond these limits, but the stan-
dard specifications may not apply and the application should be reviewed
by your AMETEK representative.
> 80 % methane
< 10 % ethane (for standard column application)
< 3 % propane
< 1.25 % total butanes
< 0.5 % C5+
< 0.6 %Water content (based on a 35 °C maximum water dewpoint tem-
perature at 120 PSIG/830 KPAG)
The surrounding ambient temperature and sample system must be at
least 5 °C (9 °F) degrees above the highest expected dewpoint tempera-
ture.
Maximum Sample Gas Temperature
0 °C to 50 °C (with Column Block)
0 °C to 165 °C (without Column Block)
Maximum Sample Cell Pressure
6.9 BARG/690 KPAG (100 PSIG)
Physical Dimensions (on Backpan)
Height: 1185 mm (46.65")
Width: 780 mm (30.7")
Depth: 254 mm (9.97"), includes backpan depth
Weight: Approximately 100 kg (220.5 lb), entire system and backpan
only (may vary, depending on system).
Options
• Other measuring ranges
• COS and methyl mercaptan measurement
• Pressure compensation
• Up to 4 V/I outputs