Speciîšżcations | 2-5
Zero Gas
Instrument zero purity Carbon Dioxide, UHP Nitrogen, or UHP Methane.
Recommended Auto-Zero interval is once per 24 hours. More frequent
Auto-Zeroing can be used to reduce temperature drift in installations with
widely varying ambient temperatures.
Ambient Limits
Temperature
From 0 °C to 50 °C (32 °F to 122 °F).
The surrounding ambient temperature must be at least 5 °C (9 °F) degrees
above the highest expected dewpoint temperature. It is also extremely im-
portant to maintain a relatively stable ambient temperature in the vicinity
of the analyzer, with no rapid temperature fluctuations.
Humidity
0–95 % RH
If ambient conditions typically exceed 70 % RH, the environmental purge
options is required.
Environmental purge set at 10 psi will consume a 300 ft
3
cylinder in
approximately 1 year.
Maximum Altitude
2000 m
Sample Pressure Requirement
Maximum Cell Pressure 1380 kPag (200 psig)
Without Filterblock Assembly:
Application dependant, refer to Final "As-Built" drawings for maximum
pressure ratings for your system.
With Filterblock Assembly:
Application dependant, refer to Final "As-Built" drawings for maximum
pressure ratings for your system.
