4-16 | Model 933S UV Analyzer
Saving Analyzer Conguration Parameters
The software allows you to save current analyzer (device) setup configu-
ration parameters to a file that can be used later as a reference or to restore
the parameters to their original settings (if they are changed). This is use-
ful to create a backup copy of the analyzer configuration parameters, or if
you will be replacing the analyzer’s EEPROM.
If major changes are made to the analyzer configuration parameters,
it is important to make a backup copy of these settings. This file can
then be used to restore all parameters to their original (or latest) set-
tings in the event of an EEPROM failure or it being damaged physi-
cally or electrically.
To save configuration parameters:
From the General tab, click Save Configuration to view the Save As
dialog box. Type (in the Windows navigation field) C:\Program Data\AM-
ETEK Configurator\S93xs\data and click Save (AMETEK does not recom-
mend saving this file to another path). The file will be saved with a “.dev
extension. This file can be used only by the Configurator Software.
Restoring Analyzer Conguration Parameters
The software allows you to retrieve previously saved analyzer configura-
tion parameters to restore the parameters to their original (or last saved)
settings. This is useful if you wish to recover analyzer configuration
parameters if changes have been made that you do not wish to keep, or to
restore the settings after replacing the analyzer’s EEPROM. The configu-
ration parameters are stored in a “.dev” file on the computer.
To restore configuration parameters:
1. From the General tab, click Restore Configuration to view the Open
dialog box. The message “Preparing to restore. Click OK to continue.
will appear. Click OK if you wish to restore configuration parameters.
2. Type (in the Windows navigation field) C:\Program Data\AMETEK
Configurator\S93xs\data, select the file, and click Open. To restore
configuration parameters to the analyzer you are communicating
with, click YES to the prompt “Would you like to save the settings to
EEPROM?”.
During the parameters restoration, the analyzer will temporarily go
OFFLINE”. After the restoration is complete, the Name and Descrip-
tion will change to reflect those of the saved parameters.
NOTE
General (tab)
<<Save Conguration>>
General (tab)
<<Restore Conguration>>