Specications | 2-7
Carbon Dioxide Sensor
Accuracy
CO
2
: ± 5.0 % full-scale of standard ranges.
Based on a full-scale of 5 % CO
2
in pipeline application, where
the sensor is in a temperature-controlled Oven.
Cross-Talk
No significant cross-talk from water, H
2
S, or hydrocarbons onto the CO
2
measurement.
Temperature Drift
1 % of customer full-scale over a 15 °C temperature change.
24-Hour Zero Drift
1 % of customer full-scale.