6-58 | Model 933S UV Analyzer
Take extreme care to avoid damaging the threads on the cable entry
glands on the Lower Electronics Enclosure and all threaded parts
on or in the Solenoid Junction Box, Solenoid Block, Heated Column
Block, and optional Display assemblies. Clean, defect-free threads are
essential to ensure a flameproof connection.
When performing equipment maintenance in hazardous areas, all
safety standards and procedures must be followed, as specified by the
Owner Company, local electrical-inspection authority, and National/
EU regulations.
Do not apply AC power to the analyzer if any of its flamepaths appear
to be scratched, dented, or worn. Applying power to an analyzer with
a damaged flamepath is dangerous and could result in serious injury
or death to personnel and/or serious damage to equipment.
Replace the parts immediately if damage or wear is apparent. Contact
AMETEK immediately if there is any doubt about the integrity of any
flamepath.
!
WARNING
!
WARNING
!
CAUTION