Maintenance and Troubleshooting | 6-79
Decimal Alarm
Bit Value Type Alarm Condition / Description and Suggested Corrective Action
5 32 Fault f: Lamp Failure
If a PMT Reference signal falls below 750 mV for 5 minutes, it indicates the lamp
has failed and requires replacement.
Corrective Action:
From the Optical Bench Data tab, check the PMT Reference value for
each lter. If any of these signals are below 750 mV, take appropriate safety
precautions and replace the Source Lamp(s) immediately (see “Source Lamp
Replacement” in this chapter).