Quick start
Easy and very fast operation of all standard functions of the device. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Page 3
Damped Osc. 3MHz Quick start
Press CHANGE and the test parameters parameter can be changed.
Select the desired parameter with the related function key and change the value by turning the front panel knob.
The cursor allows the user to define the digit to be changed (fast or slow change).
Press START and the test starts immediately with the displayed test parameters.
The operator now can navigate with the Cursor from parameter to parameter. The blinking parameter can be
changed by turning the front panel knob.
Press ESC will bring the user back to the previous menu level.
All function keys except F2 (manual trigger) can Stop the test
routine.
Burst specification as per IEC 61000-4-18
Burst duration Td 3 MHz 50 ms ±20 %
10 MHz 15 ms ±20 %
30 MHz 5 ms ±20 %
tr 300 ms
f 5000/s for 3, 10, and 30 MHz
T 1 min