EasyManua.ls Logo

Ametek OCS 500N6 Series - Technical Specifications and Data; Slow Damped Oscillatory Wave Specs (IEC 61000-4-18)

Ametek OCS 500N6 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMETEK CTS OCS 500 N6 series
Manual for Operation V 5.24 26 / 80
5.
Technical data
5.1.
Slow damped oscillatory as per IEC 61000-4-18
Test Level
Output voltage
250 V 2`500 V 10 %
250 V 3`000 V 10 % HV out on model N6.5; N6.6; N6.7; N6.8
Wave form (open circuit):
Rise time
75 ns 20 %
Oscillation frequency
100 kHz and 1 MHz 10 %
Decaying
Peak 5 must be > 50 % of peak 1 value
Peak 10 must be < 50 % of peak 1 value
No requirements for other peaks
Source impedance
Coupling capacitors
200 20 %
0,5 µF 10 %
Polarity
Positive, Negative
Repetition rate
At least 40/ s for 100 kHz and
At least 400/ s for 1 MHz
Trigger
Trigger of bursts
AUTO, MANUAL, EXTERN (min. repetition time 0.1 s)
Repetition rate, Rep
20/ s 50/ s for 100 kHz
20/ s 400/ s for 1 MHz
Burst duration td
0.1 s 99.9 s
Burst repetition time tr
0.1 s 99.9 s
Test duration T
0:01 min 99:59 min 1 s, endless
Output
Direct at the front panel
2 banana plugs for:
- For connect ext. coupling devices with firmware V 3.30axx and higher
Ri = 100 for coupling set to “Ext CN”
Ri = 200 for coupling set to\
- For previous firmware versions up to V3.24a xy
Ri =200 for coupling set to\
coupling “ExtCN” not awailable
- Magnetic Field antenna 100 Am
-1
(Firmware Version 1.13 and higher)
- Magnetic Field antenna 60 Am
-1
(Firmware Version 1.10 up to 1.12)
Coupling network 1 phase
To L, N, PE all combinations
Coupling network 3 phase
L1...L3-N, L1...L3-PE, N-PE, L1+L2+L3-PE , L1+L2+L3+N-PE
EUT Supply
AC 250 V / 16 A / 50/60 Hz
DC 250 V/10 A

Table of Contents

Related product manuals