AMETEK CTS OCS 500 N6 series
Manual for Operation V 5.24 26 / 80
Slow damped oscillatory as per IEC 61000-4-18
250 V – 2`500 V 10 %
250 V – 3`000 V 10 % HV out on model N6.5; N6.6; N6.7; N6.8
Wave form (open circuit):
Peak 5 must be > 50 % of peak 1 value
Peak 10 must be < 50 % of peak 1 value
No requirements for other peaks
Source impedance
Coupling capacitors
200 20 %
0,5 µF 10 %
At least 40/ s for 100 kHz and
At least 400/ s for 1 MHz
AUTO, MANUAL, EXTERN (min. repetition time 0.1 s)
20/ s – 50/ s for 100 kHz
20/ s – 400/ s for 1 MHz
0:01 min – 99:59 min 1 s, endless
Direct at the front panel
2 banana plugs for:
- For connect ext. coupling devices with firmware V 3.30axx and higher
Ri = 100 for coupling set to “Ext CN”
Ri = 200 for coupling set to “\”
- For previous firmware versions up to V3.24a xy
Ri =200 for coupling set to “\”
coupling “ExtCN” not awailable
- Magnetic Field antenna 100 Am
-1
(Firmware Version 1.13 and higher)
- Magnetic Field antenna 60 Am
-1
(Firmware Version 1.10 up to 1.12)
To L, N, PE all combinations
L1...L3-N, L1...L3-PE, N-PE, L1+L2+L3-PE , L1+L2+L3+N-PE
AC 250 V / 16 A / 50/60 Hz