Chapter 3 — Options Verification 3-4 Option 880, GSM/GPRS/EDGE Signal Analyzer Verification
MS2720T MM PN: 10580-00342 Rev. D 3-11
20. Record the measured “Burst Power” value into the “Displayed Burst Power” row of Table A-73, then
subtract the “Displayed Burst Power” value from the “Adjusted Output Power” reading that was
obtained in Step 19. Then record the “Calculated Burst Power Error” and the displayed value of
“Frequency Error” into Table A-73.
21. Record the displayed EVM(rms) into Table A-81, “Demodulation, At 850 MHz, –10 dBm Level,
EdgeBurst_4slot Pattern (EDGE)” on page A-40
22. Verify that the measured values in Step 20 and Step 21 are within specifications.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers: