Chapter 3 — Options Verification 3-8 Option 883, TD-LTE Signal Analyzer Verification
MS2720T MM PN: 10580-00342 Rev. D 3-33
12. Record the MS2720T Frequency Error reading into the section
“750 MHz, –30 dBm” in Table A-94.
13. Verify that the values are within specification.
14. Set the calibration factor frequency of the power sensor to 2350 MHz.
15. Set MG3700A Frequency to 2350 MHz. Press the Set key.
16. Adjust the MG3700A Level setting with the knob so that the power meter reads 12.0 dBm ± 0.5 dB.
17. Change the Center Frequency of the MS2720T to 2350 MHz.
18. Record the MS2720T Frequency Error reading into the section “2350 MHz, –12 dBm” in Table A-94.
19. Verify that the values are within specification.
20. Adjust the MG3700A Level setting to approximately –18 dBm so that the power meter reads
30.0 dBm ± 0.5 dB.
21. Record the MS2720T Frequency Error reading into the section “2350 MHz, –30 dBm” in Table A-94.
22. Verify that the values are within specification.
23. Press the Setup main menu key and set the BW to 20 MHz. Then repeat Step 5 Substep a through
Step 22 by using the TDLTE-E-TM-3-3_20M pattern for Step 5 Substep i.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals