Appendix A
Decontaminating the Sample Block
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 109
Notes
A
Cleaning the Sample Wells
IMPORTANT! Wear powder-free
gloves when you perform this
procedure.
1. Identify the contaminated wells of the sample
block (see “Troubleshooting” on page 64).
2. Remove the plate and the tray holder.
3. Close the tray. Apply pressure to the right side of
the tray and at an angle.
4. Manually raise the block from the ROI Inspector
window:
a. Click (or select FileNew).
b. In the New Document wizard,
click .
c. In the SDS software, select
InstrumentCalibrate.
d. In the warning dialog box, click to
lower the sample block.
The ROI Inspector dialog box opens.
e. In the ROI Inspector dialog box,
click .
5. Power off, then unplug the 7300/7500/7500 Fast
system. Allow it to cool for 15 min.
Press
here
>00:15:00

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals