EasyManuals Logo

Arbiter Systems 1092A User Manual

Arbiter Systems 1092A
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
86 Serial Communication and Command Set
10.2.14 Out-of-Lock Commands
Set Out-of-Lock Time
Command: (-)nnK
(-)nnK configures the amount of delay time (in minutes) following loss of satellite synchronization
before an out–of–lock signal is generated and output via rear panel connector (if Option 93 is
installed). A negative number turns the out-of-lock function OFF. n = 0 sets the out-of-lock time
to zero delay. n = 1 to 99 sets the amount of delay time (in minutes).
Response:
C
10.2.15 Miscellaneous Commands
Return Firmware Version
Command: V
V returns the firmware version date of the installed ROM.
Response: dd mmm yyyy
C
Return Display Buffer
Command: Z
Z returns the contents of Display Buffer.
Response: Echoes current display (40 characters); no line wrap. For display contents, see Chap-
ter 6, Startup and Basic Operation.
Set Option Control
Command: m,n,k,lXI
m,n,k,lXI configures the specified option in the clock, where m = 0 for the main board and m = 1
for the auxiliary board.
Where:
m = 0 for Main board option, and m = 1 for Aux board option.
n = option number (see Table 10.9 below)
k = security key; 1092 or 1093
l = use only if n specifies Option 28. l = 0 for 60 Hz and l = 1 for 50 Hz.
Response:
C
Main Board Option, m=0 none 19
Aux Board Option, m=1 none 03 20A 27 28 29 32 33 34 35
Option Number, n= 0 1 2 3 4 5 6 7 8 9
Table 10.9: Option Control Settings

Table of Contents

Other manuals for Arbiter Systems 1092A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arbiter Systems 1092A and is the answer not in the manual?

Arbiter Systems 1092A Specifications

General IconGeneral
BrandArbiter Systems
Model1092A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals