EasyManuals Logo

Arbiter Systems 1092A User Manual

Arbiter Systems 1092A
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
60 Relay Contacts and Event Inputs
9.2.6 Firmware Setup
Reconfiguration of the firmware may also be required to allow measurement and display of Event
Time Data and/or 1 PPS Deviation. See Figure 7.17 for detail on configuring the Event/Deviation
parameters from the front panel. See Section 10.2.4 for details on using the RS-232 interface.
9.2.7 Displaying Data
Event and deviation data can be accessed from either the front panel or via RS-232 com-
mands. The following paragraphs describe the steps required to access data using the front panel
EVENT/DEVIATION key. When pressing the EVENT/DEVIATION key, the display will enter a
circular scroll, which begins by showing the data (if any is present) for Channel A, as previously
configured for Event Recording in Section 10.2.4. The readout will display one of the event times
using the following format:
CH A EVENT #nnn TIME
ddd:hh:mm:ss.sssssss
Where:
nnn = event number (001 to 400)
ddd = day of year of the event (1 to 366)
hh = hour of the event (00 to 23)
mm = minute of the event (00 to 59)
ss.sssssss = second and fractional seconds of the event (e.g. 59.9999999)
Pressing the UP and DOWN keys will scroll the display through all events presently stored in
the event time buffer. If the event display mode is exited and then re-entered, the first event data
displayed for a given channel will correspond to the same event number as was last displayed for
that channel. However, the data itself may be changed if it has been overwritten.
To clear the event buffer, press the EVENT/DEVIATION key again, while viewing event data.
This will cause the display to issue the following prompt:
CLEAR EVENT (A)?
NOTE: If the event capture channel is not configured for event time recording, the CLEAR EVENT
prompt will not be displayed.
Clocks display the CLEAR EVENT prompt only if the event capture channel was previously
configured for event time recording. Press ENTER during this display to delete all of the records
in the event buffer, and to allow recording of new event times, starting with event number 001.
If the event capture channel is configured for 1–PPS Deviation (via Event/Deviation Setup
Menu), the readout will display the deviation of the 1–PPS input signal. In this case, the readout
display will have the format:
A 1 PPS XXXXX.XX ?S
SIGMA: XXXXX.XX ?S
Where: The top number is the mean (average) value of the most recent 16 records in the event
buffer, and represents the mean deviation (in microseconds) of the measured 1–PPS signal from

Table of Contents

Other manuals for Arbiter Systems 1092A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arbiter Systems 1092A and is the answer not in the manual?

Arbiter Systems 1092A Specifications

General IconGeneral
BrandArbiter Systems
Model1092A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals