Physical Interface AW00123409000
104 Basler ace USB 3.0
Line Status for Opto-isolated Input Line (Line 1)
Line Status for Direct-coupled GPIO Lines (Line 3, Line 4), Set for Input
External Conditions Resulting Status
Line Inverter Status Electrical Status Logical Line Status
Parameter Value
Binary Expression
Disabled Input Open or
Connection at z Status
False 0
Voltage Level Low False 0
Voltage Level High True 1
Enabled Input Open or
Connection at z Status
True 1
Voltage Level Low True 1
Voltage Level High False 0
Table 28: Line Status for Different External Conditions: Line 1 (Opto-isolated Input)
External Conditions Resulting Status
Line Inverter Status Electrical Status Logical Line Status
Parameter Value
Binary Expression
Disabled Input Open or
Connection at z Status
True 1
Voltage Level Low False 0
Voltage Level High True 1
Enabled Input Open or
Connection at z Status
False 0
Voltage Level Low True 1
Voltage Level High False 0
Table 29: Line Status for Different External Conditions: Line 3, Line 4 (Direct-coupled GPIO Input)