AW00123409000 Image Acquisition Control
Basler ace USB 3.0 167
For more information about event notification, see Section 7.17 on page 314.
The frame burst trigger wait signal will only be available when hardware frame
burst start triggering is enabled.
Time
= Camera is in a "waiting for
frame burst start trigger" status
Frame Burst
Trigger Wait
Signal
Frame Acquisition
Exp.
Sensor
Readout
Frame Acquisition
Exp.
Frame Acquisition
Exp.
Frame Acquisition
Exp.
Frame Acquisition
Exp.
Frame Acquisition
Exp.
Frame Burst
Trig Signal
Fig. 75: Frame Burst Trigger Wait Signal