Error
Code
Error Message Possible Cause Error Resolution
E067
E072
E074
Camera infeed:
receive timeout (dish
not detected)
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
E077
Camera infeed:
cylinder not in
position
E078
Camera infeed:
cylinder not in rest
position
E079
Camera infeed:
rotation motor current
limit overload
E110
Camera: cylinder not
in position
E111
Camera: cylinder not
in rest position
E112
Camera: no dish lid
vacuum detected
Possible obstruction
between suction cup
and plate lid.
1. Check and clean the suction
cup.
2. Check for broken or missing
lid and replace it with a new
one.
3. Select Reset Error [F5].
4. Select Start [F2] to restart the
module.
E113
Camera: dish lid
vacuum loss detected
E116
Camera: no rotate
disc detected
Various causes.
1. Check the rotate disc.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
E121
Camera: receive
timeout (dish not
detected)
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
E122
Camera: possible
sticky dish at stopper
door
Possibly a plate has
become stuck at the
stopper door.
1. Remove the plate from the
camera stopper.
2. Add the plate to a Destacker to
be processed by the Track
again.
3. Select Reset Error [F5].
4. Select Start [F2] to restart the
module.
65