Error
Code
Error Message Possible Cause Error Resolution
E124
Camera: camera
ready to image, event
timeout
Various causes.
1. Select Reset Error [F5].
2. Select the X icon in the top
right corner of the ReadA
application title bar to close
the application.
3.
Select the shortcut on the
desktop to re-launch the
ReadA application and sign in.
4. Select Start [F2] to restart the
module.
E150
Camera outfeed:
possible sticky dish at
stopper door
Possibly a plate has
become stuck at the
stopper door.
1. Remove the plate from the
camera outfeed stopper.
2. Add the plate to a Destacker to
be processed by the Track
again.
3. Select Reset Error [F5].
4. Select Start [F2] to restart the
module.
E151
E155
Camera outfeed:
receive timeout (dish
not detected)
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
E158
Camera outfeed:
cylinder not in
position
E159
Camera outfeed:
cylinder not in rest
position
E160
Camera outfeed:
rotation motor current
limit overload
E161
Camera outfeed:
vacuum not detected
on scanner
The scanner lifting
cylinder may be dirty.
1. Check and clean the scanner
lifting cylinder.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
E162
Camera outfeed:
conveyor motor
current limit overload
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.