10
Beam
®
, Operating Manual, Rev. 4.6.1 01/2016, GmbH
Directions for use
2
Pay attention to the maximum operating period of used single-use electrodes
and inform your patient about the period and the affixing of new electrodes.
Only use single-use electrodes which are appropriate for the skin type of your
patient.
2.6 Important Notes for the Patient
Please comply with the instructions of your physician.
Stay calm during the reading until the process is finished.
The ECG Recorder Beam
®
comes with a cable recognition. Wait for a short
audible signal on plugging in the cable, only then is the cable correctly
plugged in. For the correct positioning you can find a red dot on the plug
and on the case, which belong together. Handle the connection with care -
the plug should slide without too much pressure into the port.
In the normal operating mode the ECG Recorder flashes green every 2
Seconds. If an error occurs the recorder will issue audible signals.
The following errors can be remedied by you, if this was discussed
previously with the physician:
The Recorder flashes red every 2 seconds:
The battery is flat – replace the battery.
The Recorder flashes red twice every 5 seconds:
One electrode has insufficient contact – check the positioning of the
electrode as shown by the physician.
The following errors can only be remedied with the assistance of the phy-
sician:
The recorder flashes red three times every 6 seconds:
The recording memory is full – Contact your physician for further
instructions.
The recorder flashes red without interruption:
Remove the device and contact your physician for further instructions.
Warnings:
The ECG recorder Beam
®
should not come in contact with any fluids. Do
not wear the device when taking a shower or bath!
If water should penetrate the device it can not be used anymore. Please
remove the battery immediately.
If skin irritations, rashes, or itching occurs in the area of the electrodes
please remove the electrodes and inform your physician.
Inform your physician if the device does not flash green every 2 seconds.
Please find a template copy for the patient in the attachment!