PN 177196BB
7-20
DATA MANAGEMENT AND REVIEW
DATA ANALYSIS REVIEW - WBC PARAMETERS
Summary
1. Table 7.6 summarizes the WBC parameters generated by the A
C
•T 5diff instrument and
how each is derived.
2. For details on the methodologies used, review
Heading 2.2 MEASUREMENT PRINCIPLES and
Heading 2.7 PARAMETER DEVELOPMENT, in the Online Help System or the Instructions for
Use manual
.
3. For details on the flags and interpretive messages for that can be generated, review
Heading 9.8 FLAGS AND MESSAGES GENERATED BY THE INSTRUMENT, in the Online Help
System or the Instructions for Use manual
.
Table 7.6 WBC Parameter Summary
Category Parameter Source of Data
Directly Measured WBC (White Blood Count) Differential lysis plus Coulter Principle
BA % (Basophil Percentage) Differential lysis plus Coulter Principle
Derived from the DiffPlot NE % (Neutrophil Percentage)
A
C
V Technology
LY % (Lymphocyte Percentage)
A
C
V Technology
MO % (Monocyte Percentage)
A
C
V Technology
EO % (Eosinophil Percentage)
A
C
V Technology
Computed NE # (Neutrophil Absolute Number)
LY # (Lymphocyte Absolute Number)
MO # (Monocyte Absolute Number)
EO # (Eosinophil Absolute Number)
BA # (Basophil Absolute Number)
Neut% WBC×
100
-----------------------------------
Lymph% WBC×
100
----------------------------------------
Mono% WBC×
100
-------------------------------------
Eos% WBC×
100
--------------------------------
Baso% WBC×
100
-----------------------------------