CB (E7) 06/2018 Page 136/174
To maintain the proper function of the CO
2
sensor, never spray the sensor with any cleaning detergents
or disinfectants. Generally perform any cleaning only when the chamber is turned off.
CAUTION
Damage to the CO
2
sensor.
Do NOT spray cleaning detergents directly on the CO
2
sensor.
Wipe the sensor with a lint-free cloth soaked with cleaning agent.
Use only the detergents or disinfectants recommended by BINDER. Other products are not approved,
since they could in particular damage the CO
2
sensor.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove any cleaning agents from the surfaces by using a moistened
towel. Let the chamber dry.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
With every decontamination method, always use adequate personal safety controls.
Following cleaning, leave the chamber door open or remove the access port plugs.
The neutral cleaning agent may cause health problems in contact with skin and if ingested.
Follow the operating instructions and safety hints labeled on the bottle of the neutral
cleaning agent.
Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective
gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes.
CAUTION
Contact with skin, ingestion.
Skin and eye damage due to chemical burns.
Do not ingest. Keep away from food and beverages.
Do NOT empty into drains.
Wear protective gloves and goggles.
Avoid skin contact.
Following use of the neutral cleaning agent and prior to hot-air sterilization, remove any agent
residues by using a moistened towel in order to avoid formation of permanent residues.