CB (E7) 06/2018 Page 161/174
Chamber size
56
170
260
Electrical data
IP system of protection acc. to EN 60529
IP
20
20
20
Nominal voltage
(+/-10 %)
at 50 Hz power frequency
V
200-230
200-230
200-230
at 60 Hz power frequency
V
200-230
200-230
200-230
Current type
1N~
1N~
1N~
Nominal power
kW
***
1.40
***
Power plug (IEC connector plug)
Grounded plug IEC 7/7
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Chamber fuse 4x20mm / 250V / time-lag
(T), external
Amp 10 10 10
Additional temperature fuse class 1 (DIN
12880), internal
°C 240 240 240
Different electrical data for CB-UL constructed for the USA and Canada
Nominal voltage
(±10 %)
at 50 Hz power frequency
V
100-120
100-120
100-120
at 60 Hz power frequency
V
100-120
100-120
100-120
Current type
1N~
1N~
1N~
Power plug
NEMA
5-15P
5-20P
5-20P
Chamber fuse 6.3x32 mm / 250V / time-lag
(T), external
Amp 12.5 12.5 12.5
Environment-specific data
Energy consumption at 37 °C/ 98.6 °F
Wh/h
***
60
***
*** Data not yet determined
Note: The recovery times of the gas concentrations inside the chamber after the door is opened coincide
with a connection pressure of 2.0 bar / 29 psi. Decreasing supply pressure results in longer recovery
times.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of 22 +/- 3 °C / 71.6 +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
Chamber with active humidification: Bringing a source of humidity into the inner chamber will
affect the minimum humidity specification
25.5 Equipment and Options (extract)
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
Regular equipment
Multifunction touchscreen controller MB2 for temperature, CO
2
, O
2
(chamber with O
2
control) and
humidity (chamber with active humidification)
CO
2
infra-red absorption measuring system
Silicone jacket heating
Hot-air auto sterilization at 187.5 °C / 369.5 °F
Gas mixing head