CB (E7) 06/2018 Page 156/174
24.5 Sterilization
The sterilization is running.
Alarm message “Sterilization”
Sterilization cycle is running
Acknowledge the alarm on the
controller. Do not open the door.
24.6 Humidity (chamber without active humidification)
Condensations inside the
chamber.
Permadry™ pan filled with water
when the chamber is not
Empty Permadry™ pan when the
chamber is not operating.
Doors not closed properly.
Reduce humidity (chap. 6.9)
Condensation on the doors or
on the optional divided inner
door, gas-proof.
Chamber placed on very cold
floor.
Place the chamber on a BINDER
stand to increase the distance to the
Doors not closed properly.
Reduce humidity (chap. 6.9)
Door heating set too low.
Contact BINDER Service to adjust
Too low humidity inside.
Increase humidity (chap. 6.9)
Chamber door not properly
closed.
Completely close chamber door.
Permadry™ pan empty.
Fill the outer Permadry™ pan with
water up to the marking on the inner
pan with distilled, sterile water. The
pan must have thorough contact to
the bottom of the inner chamber,
24.7 Humidity (chamber with active humidification)
Control accuracy of +/- 3 % r.h.
Door opened very frequently. Open doors less frequently.
Humidity fluctuation, together
with temperature fluctuation > 1
°C with a set-point approx. 3 °C
above ambient temperature.
Place of installation too hot.
Select cooler place of installation or
contact BINDER service.