EasyManua.ls Logo

Binder CB 170 - Permadry Humidity Control Adjustment

Binder CB 170
175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CB (E7) 06/2018 Page 56/174
6.9 Humidity control of the Permadry™ system
With the chamber with active humidification, humidity setting via the Permadry System is not required.
The Permadry™ system with its 2-pan water system guarantees a maximum humidity of up to 95 % r.H.
in the inner chamber, which remains condensation-free. This performance assumes an average ambient
temperature of 22 +/- 3 °C / 71.6 +/- 5.4 °F and a working temperature in the inner chamber of 37 °C /
98.6 °F.
You can increase or decrease humidity slightly via the heating of the water pan if required.
Humidity is infinitely adjustable in whole numbers from -5 (low) to +5 (high).
Factory setting: “0” (medium).
Setting:
Path: Main menu > Settings > Various
Submenu “Various”.
Select the field “Humidity control
“Humidity control” entry menu. Enter the desired
humidity and confirm entry with Confirm icon.
If the ambient temperature deviates by more than +/- 5 °C from the values recommended by
the manufacturer, the conditions for maximum air humidity with condensation-free inner
chamber are no longer guaranteed. Contact BINDER Service for assistance.