CB (E7) 06/2018 Page 155/174
2
Having reached the setpoint,
CO
2
falls below the setpoint by
more than the set tolerance
range value (more than 10
min.) or CO
2
doesn’t reach the
tolerance range within 3 hours
from turning on the chamber or
closing the door.
Alarm message “CO
2
range”
Temporary disturbance of the
CO
2
control.
Acknowledge the alarm on the
controller. Turn off the chamber. After
approx. 5 minutes turn on the chamber
again. Upon renewed alarm, contact
CO
2
sensor defective.
Acknowledge the alarm on the
controller. Contact BINDER Service.
2
Too low O
2
concentration.
Having reached the setpoint,
O
2
falls below the setpoint by
more than the set tolerance
range value (more than 10
min.) or O
2
doesn’t reach the
tolerance range within 3 hours
from turning on the chamber or
closing the door.
Alarm message “O
2
range”.
Temporary disturbance of the
O
2
control.
Acknowledge the alarm on the
controller. Turn off the chamber. After
approx. 5 minutes turn on the chamber
again. Upon renewed alarm, contact
O
2
sensor defective.
Acknowledge the alarm on the
controller. Contact BINDER Service.
Too low gas concentration.
The concentration of CO
2
or
O
2
(chamber with O
2
control)
does not reach the adjusted
set value.
Doors not closed properly.
Connected gas cylinder is
empty or not opened.
Open or replace gas cylinder.
Gas cylinder is not connected
correctly.
Correctly connect the gas cylinder.
Gas hose is dirty or
obstructed.
Check the tube system for dirt
accumulation or obstruction, clean or
Recovery time (up to 5 vol.-%
CO
2
) after doors were open for
2 minutes is > 10 minutes.
Check gas supply (cylinder,
connections, hose system).
2
Unusually high gas
consumption.
Contact BINDER Service
Gas fine filter not connected
correctly.
The displayed actual value of
CO
2
deviates largely compared
with a reference method.. The
pH indicator of the cell culture
medium changes its normal
CO
2
control not adjusted. Calibrate and adjust CO
2
control.
CO
2
sensor defective.
Transfer the cultures to another
incubator and contact BINDER Service.
Chamber with O
2
control:
The displayed actual value of
O
2
(option) deviates largely
compared with a reference
Calibrate and adjust O
2
control.
O
2
sensor defective.
Transfer the cultures to another
incubator and contact BINDER Service.
No gas control after hot-air
sterilization.
Sterilization cycle still running.
Terminate hot-air sterilization.
Information message not
confirmed.
Confirm the information message with
“Ok”.
Inner chamber temperature
above 80 °C
Let the inner chamber temperature
decrease below 80 °C.