Rev 3.1 9. IBP 155
BM5 (CS,CX) Users Manual
SCALE (Setting size of measurement waveform)
You can set the pressure range for measurement waveform on this menu.
The selectable values mean the maximum blood pressure range value that can be shown in a
waveform.
SETTINGZERO
CHANGE
NAME
:ART
SCALE:
160
ALARM
MAIN
MENU
300 80
200 60
> 160 30
100
SCALE:
160
MAIN
MENU
Alarming Limits for ART
Alarming limits vary according to measuring positions.
The settable alarming range for systolic pressure, diastolic pressure and mean pressure is - 50 ~
350mmHg.
IBP1 ALARM LIMIT & LEVEL
UNITS LOW LEVEL
ART-S
ART-M mmHg
150
50 115
RETURN
HIGH
MEDIUM
MEDIUM
ALARM
ON/OFF
ON
OFFmmHg 70
ART-D mmHg 40 100 MEDIUMON
ART-PR BPM 50 150 MEDIUMON
SETTINGZERO
CHANGE
NAME
:ART
SCALE:
160
ALARM
MAIN
MENU
MAIN
MENU
CHANGE
NAME
ART
ART
SCALE:
30