Rev 3.1 9. IBP 157
BM5 (CS,CX) Users Manual
IBP SETTING (Setting Various Functions)
Other menus are to be applied for special functions to process pressure signals in various ways.
SETTINGZERO
CHANGE
NAME
:ART
SCALE:
160
ALARM
MAIN
MENU
BP
FILTER:
OFF
PREV
MENU
MAIN
MENU
Setting three labels of ART, FEM and UAP displaying PULSE-RATE among labels, the functions of
PULSE-RATE DISPLAY and DISCONNECT ALARM will be added.
BP
FILTER:
OFF
PULSE
RATE:
OFF
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU
BP FILTER: It filters waveforms by selecting three frequency bands.
OFF 0Hz ~ 40Hz
12Hz 0Hz ~ 12Hz Generally recommended for monitoring
20Hz 0Hz ~ 20Hz Used for processing waveform components of higher
frequency. Pressure value can be increased with this filter.
BP
FILTER:
OFF
PULSE
RATE:
OFF
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU