Rev 3.1 9. IBP 158
BM5 (CS,CX) Users Manual
> OFF
12 Hz
20Hz
BP
FILTER:
OFF
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU
PULSE RATE: Setting display of blood pressure pulse rate.
BP
FILTER:
OFF
PULSE
RATE:
OFF
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU
BP
FILTER:
OFF
PULSE
RATE:
ON
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU
CAL. TRANSDUC: A function to adjust a Transducer error on the monitor
A function to adjust an error value based on the other index manometer.
How to Adjust
1. Select a menu by pressing the knob switch key.
2. Measure blood pressure along with another index manometer.
3. Compare the measured values of ‘mmHg’ for both manometers.
4. Adjust the error value on the parameter menu screen by turning knob switch.
5. Terminate the menu by pressing the knob switch key again.
DISCONN ALARM: (Alarming function for disconnection)
DISCONN ALARM MENU will be displayed when measurement label is set for ART, FEM and UAP.
This function will be activated upon the following two conditions.
1. In case MEAN PRESSURE is not higher than 25mmHg.
2. In case the Disconnect Alarm is set ‘ON’.