Rev 3.1 9. IBP 159
BM5 (CS,CX) Users Manual
Midium alarming sound will be generated when the DISSCONNECTED ALARM is activated, and the
alarming message “DISCONNECTED” will be displayed on the parameter screen.
BP
FILTER:
OFF
PULSE
RATE:
OFF
DISCONN.
ALARM:
OFF
PREV
MENU
MAIN
MENU
BP
FILTER:
OFF
PULSE
RATE:
OFF
DISCONN.
ALARM:
ON
PREV
MENU
MAIN
MENU
DISCONNECTED
200
80
mmHg
ART
S
Troubleshootings for a case the measured value is different from the expected value
Description
Action to Take
In case there are air bubbles in tubes
Remove the air bubbles
In case an extension tube is connected
Remove the extension tube
In case of using blood pressure transducer
with a different sensitivity
Check position of transducer
For other cases
Perform zero-point adjustment