Rev. 3.1 1.BASIC
20
BM5 (CS,CX) Users Manual
Manufacturer’s declaration - electromagnetic immunity
The BM5 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the BM5 system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic dis
charge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood, co
ncrete or ceramic tile. If f
loors are covered with sy
nthetic material, the relati
ve humidity should be at l
east 30 %
Electrical fast
Transient / burs
t
IEC 61000-4-4
2kV for power supply lin
es 1kV for input/output li
nes
2kV for power supply li
nes
1kV for input/output lin
es
Mains power quality shoul
d be that of a typical com
mercial or hospital environ
ment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality shoul
d be that of a typical com
mercial or hospital environ
ment.
Power frequenc
y
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical comm
ercial or hospital environm
ent.