Rev. 3.1 1.BASIC
24
BM5 (CS,CX) Users Manual
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range app
lies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affec
ted by absorption and reflection from structures, objects, and people.
Immunity and Compliance Level
Immunity test
IEC 60601 Test Level
Actual Immunity Level
Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m, 80 MHz to 2.5 GHz
3 V/m, 80 MHz to 2.5 GHz
3 V/m, 80 MHz to 2.5 GHz
Guidance and manufacturers declaration - electromagnetic immunity
The BM5 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the BM5 system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -guidance
Conducted RF
IEC 61000-4-
6
3 Vrms
150 kHz to 80M
Hz
3 Vrms
150 kHz to 80 M
Hz
BM5 system must be used only in a shi
elded location with a minimum RF shiel
ding effectiveness and, for each cable t
hat enters the shielded location with a
minimum RF shielding effectiveness an
d, for each cable that enters the shield
ed location