Rev. 3.1 1.BASIC
BM5 (CS,CX) User’s Manual
Radiated RF
IEC 61000-4-
3
3 V/m
80.0 MHz to 2.5
GHz
3 V/m
80.0 MHz to 2.5
GHz
Field strengths outside the shielded loc
ation from fixed RF transmitters, as det
ermined by an electromagnetic site surv
ey, should be less than 3V/m.a
Interference may occur in the vicinity o
f equipment marked with the following
symbol:
Note 1) These guidelines may not apply in all situations. Electromagnetic propagation is affec
ted by absorption and reflection from structures, objects and people.
Note 2) It is essential that the actual shielding effectiveness and filter attenuation of the shie
lded location be verified to assure that they meet the minimum specification.
a- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadc
ast cannot be predicted theoretically with accuracy. To assess the electromagnetic environme
nt due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength outside the shielded location in which the EUT is used exceeds 3V/
m, the EUT should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocati
ng the EUT or using a shielded location with a higher RF shielding effectiveness and filter at
tenuation.
For Type A Professional ME Equipment intended for use in domestic establishment instructions
for use includes a warning:
This ME equipment is intended for use by professional healthcare personnel only.